RAPIDPOINT 500 10492730

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-03-15 for RAPIDPOINT 500 10492730 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[40390941] Siemens customer product support investigated the event and concluded following: the customer was unable to provide sample times or id's but stated they were run sometime on the (b)(6), before 14:51 when the complaint was received by siemens. From the provided files, the attempt to match the provided information was made. Data is inconclusive from the data provided as to which samples were truly suspect. Moreover, instrument (b)(4) experienced numerous occurrences of benzalkonium exposure during use life. This is a known interfering substance for the sodium sensor as listed in the rp500 operators manual.
Patient Sequence No: 1, Text Type: N, H10


[40390942] Customer reported that pco2 result did not correlate with their other rp500. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1217157-2016-00029
MDR Report Key5501570
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-03-15
Date of Report2016-03-15
Date of Event2016-02-16
Date Mfgr Received2016-02-16
Date Added to Maude2016-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRAPIDPOINT 500
Generic NameRP 500
Product CodeGKR
Date Received2016-03-15
Catalog Number10492730
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-15