LACTATE REAGENTS 10325776

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-15 for LACTATE REAGENTS 10325776 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[40671090] Siemens healthcare diagnostics has become aware of n- acetylcysteine (nac) and metamizole (dipyrone) interference with trinder and trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving n- acetylcysteine (nac) or metamizole. The instructions for use (ifu) for the advia chemistry assays cholesterol_2, cholesterol_c, enzymatic creatinine_2, glucose oxidase, triglycerides_2, tricglycerides_2 concentrated, uric acid, uric acid, concentrated reagents, and fructosamine, (in the limitations of the procedure section) will be updated to indicate that: venipuncture should occur prior to n-acetyl cysteine (nac) or metamizole administration due to the potential for falsely depressed results. The instructions for use (ifu) for the advia chemistry assays ldl cholesterol direct and direct hdl cholesterol, (in the limitations of the procedure section) will be updated to indicate that: venipuncture should occur prior to metamizole administration due to the potential for falsely depressed results. The instructions for use (ifu) for the advia chemistry assays lactate, glucose oxidase concentrated reagents (in the limitations of the procedure section) will be updated to indicate that: venipuncture should occur prior to n-acetyl cysteine (nac) administration due to the potential for falsely depressed results. An urgent field safety notice (ufsn) chc16-01. A. Ous was sent to ous customers and a urgent medical device correction (umdc) chc16-01. A. Us was sent to us customers in march of 2016. The ufsn and umdr will notify customers that venipuncture should occur before drug administration of nac or metamizole, due to the potential for depressed results as indicated above for each assay.
Patient Sequence No: 1, Text Type: N, H10


[40671091] The customer would like information on specific interferants like n-acetylcysteine (nac) and metamizole in reagents that use the trinder reaction on the advia chemistry instruments. The customer had received a letter from another manufacturer indicating possible interferences with n-acetylcysteine (nac) and metamizole and potential impact on assays that use the trinder reaction. The trinder reaction is a reaction where hydrogen peroxide is formed and subsequently reacts with a phenol derivative and aminoantipyrine in the presence of peroxidase to form a colored quinone product. Nac is the accepted antidote for acetaminophen toxicity and is justified in patients at significant risk for hepatotoxicity. Metamizole is an anti-inflammatory anti-pyretic drug banned in most countries because of the potential for nephrotoxicity. It is unknown if any patient results were reported to the physician(s). There were no reports of patient intervention or adverse health consequences.
Patient Sequence No: 1, Text Type: D, B5


[45263781] The initial mdr 2432235-2016-00131 was filed on 3/15/2016. Additional information (3/22/2016): please note additional information in report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2016-00131
MDR Report Key5502648
Date Received2016-03-15
Date of Report2016-03-14
Date of Event2015-07-17
Date Mfgr Received2016-03-22
Date Added to Maude2016-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAARTI AZIZ
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242683
Manufacturer G1RANDOX LABORATORIES LTD. (REG# 8020890)
Manufacturer Street55 DIAMON ROAD CRUMLIN
Manufacturer CityCO. ANTRIM, BT294QY
Manufacturer CountryUK
Manufacturer Postal CodeBT29 4QY
Single Use3
Previous Use Code3
Removal Correction Number2517506-03/22/2016-001-C
Event Type3
Type of Report3

Device Details

Brand NameLACTATE REAGENTS
Generic NameLACTATE REAGENTS
Product CodeCFJ
Date Received2016-03-15
Model NumberLACTATE REAGENTS
Catalog Number10325776
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.