HELIOS 3000 TRACK MOUNTED LIGHT HL3T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-17 for HELIOS 3000 TRACK MOUNTED LIGHT HL3T manufactured by Pelton & Crane.

Event Text Entries

[40853977] Upon evaluation by the dr. It was determined the roll pins and set screws were not installed by the distributor during installation. The roll pins and set screws will prevent the light from unscrewing from the trolly after installation. The pelton & crane installation instructions clearly states to properly install the roll pins and set screws during installation of the track light. The installation instructions also lists warnings to ensure the roll pins and set screws are installed. Pelton & crane recommended to the dr. To call his local distributor to re-install the light to the trolley but the dr. Declined to do so.
Patient Sequence No: 1, Text Type: N, H10


[40853978] It was reported from a doctors office that a pelton & crane helios track mounted dental light had unthreaded from the trolley and fell to the floor. There was no patient involved and no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2016-00006
MDR Report Key5507154
Date Received2016-03-17
Date of Report2016-02-22
Date of Event2016-02-22
Date Mfgr Received2016-02-22
Device Manufacturer Date2010-02-17
Date Added to Maude2016-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHELIOS 3000 TRACK MOUNTED LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2016-03-17
Model NumberHL3T
Catalog NumberHL3T
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-17

© 2026 FDA.report