[40772809] The vest currently provides instructions for emergency access to the chest as shown in the photo attached. Further details will be gathered from the hospital and we will provide a follow up report.
Patient Sequence No: 1, Text Type: N, H10


[40772810] Patient was wearing halo vest and required emergency cpr. The staff did not know the correct procedure to remove the vest to administer cpr. The patient died.
Patient Sequence No: 1, Text Type: D, B5


[48831341] The vest currently provides instructions for emergency access to the chest as shown in the photo attached. We have initiated a preventative action to maximize the visibility of the emergency label and/or change the labels to symbols rather than text. -
Patient Sequence No: 1, Text Type: N, H10


[48831342] Patient was wearing halo vest and required emergency cpr. The staff did not know the correct procedure to remove the vest to administer cpr. The patient died.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836248-2016-00002
MDR Report Key5513714
Report SourceFOREIGN
Date Received2016-03-21
Date of Report2016-03-21
Date Mfgr Received2016-03-16
Date Added to Maude2016-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. KAREN MONTES
Manufacturer Street27051 TOWNE CENTRE DRIVE
Manufacturer CityFOOTHILL RANCH CA 92610
Manufacturer CountryUS
Manufacturer Postal92610
Manufacturer Phone9493823741
Manufacturer G1OSSUR AMERICAS, INC.
Manufacturer Street910 BURSTEIN DRIVE
Manufacturer CityALBION MI 49224
Manufacturer CountryUS
Manufacturer Postal Code49224
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRESOLVE HALO VEST SHORT LAMSWOOL
Generic NameCOMPONENT, TRACTION, NON-INVASIVE
Product CodeKQZ
Date Received2016-03-21
Model Number545700L
Catalog Number545700L
OperatorPARAMEDIC
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOSSUR AMERICAS, INC.
Manufacturer Address910 BURSTEIN DRIVE ALBION MI 49224 US 49224


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2016-03-21