STRATUS CS 10453531

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-22 for STRATUS CS 10453531 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[41172927] Siemens customer service engineer (cse) inspected the analyzer and found lamp connector melted into lamp manager. He replaced defective lamp and lamp manager. Cse calibrated standard tab, performed alignments and ran system check, everything was working as intended. Customer ran qc and it was running ok. Instrument is fully operational.
Patient Sequence No: 1, Text Type: N, H10


[41172928] Customer observed smoke coming out of the stratus chassis when running a blood sample hence turned off the stratus power. There was no report of injury due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1217157-2016-00033
MDR Report Key5516238
Date Received2016-03-22
Date of Report2016-03-22
Date of Event2016-02-24
Date Mfgr Received2016-02-24
Date Added to Maude2016-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY
Manufacturer CityNEWARK DE 19702
Manufacturer CountryUS
Manufacturer Postal Code19702
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTRATUS CS
Generic NameSTRATUS CS
Product CodeDAP
Date Received2016-03-22
Catalog Number10453531
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-22

© 2026 FDA.report