TRANSITION GATE BRACKET MOUNT FOR REG. TRACK 363980

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-22 for TRANSITION GATE BRACKET MOUNT FOR REG. TRACK 363980 manufactured by Prism Medical.

Event Text Entries

[41174327] The transition gate parts have been requested back to evaluate. An investigation will be completed and a final report submitted after completion of investigation.
Patient Sequence No: 1, Text Type: N, H10


[41174328] Motor fell out of track when health care worker pushed it away after use. No one was injured.
Patient Sequence No: 1, Text Type: D, B5


[45775208] Root cause: primary: indications of poor bracket positioning, increased deflection of the system, and excess wear on the brackets suggest that the installation is the primary cause. These issues prevented the guide screw from completely engaging with the ramp. Secondary: as the pin did not slide easily through the plastic linear bearings, the pin did not drop down as designed. Remaining in the "up" position, it allowed the lift to roll out of the track. Correction: check the installation to ensure it is level and does not have large gaps between the track. The transgate components returned should be replaced. Corrective action: primary: the installation of the remaining systems installed at the facility should be checked to ensure they are level and do not have large gaps between the track. The installation crew and facility staff should be retrained to confirm clear understanding of proper install/use of the gantry system. Secondary: the plastic bearing for the transgate was upgraded in mid-2015 (ecr 802) and current production utilizes a different metal roller bearing instead. The replaced transgate components will have metal roller bearings installed which should allow the drop pin to move more freely.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007802293-2016-00053
MDR Report Key5518223
Date Received2016-03-22
Date of Report2016-05-23
Date of Event2016-02-26
Date Mfgr Received2016-02-26
Device Manufacturer Date2014-02-28
Date Added to Maude2016-03-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVE KILBURN
Manufacturer Street10888 METRO COURT
Manufacturer CityMARYLAND HEIGHTS MO 63043
Manufacturer CountryUS
Manufacturer Postal63043
Manufacturer Phone3142198614
Manufacturer G1PRISM MEDICAL
Manufacturer Street10888 METRO COURT
Manufacturer CityMARYLAND HEIGHTS MO 63043
Manufacturer CountryUS
Manufacturer Postal Code63043
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSITION GATE BRACKET MOUNT FOR REG. TRACK
Generic NameTRANSITION GATE
Product CodeFMR
Date Received2016-03-22
Model Number363980
Device Availability*
Device Age2 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPRISM MEDICAL
Manufacturer Address10888 METRO COURT MARYLAND HEIGHTS MO 63043 US 63043


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-22

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.