STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 22 777404-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-25 for STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 22 777404-101 manufactured by Carefusion, Inc.

Event Text Entries

[41419862] (b)(4). If more information becomes available a follow up will be submitted. (b)(4). The carefusion field service engineer (fse) was onsite and stated the arm was tight so when the customer adjusted the arm, it probably put pressure and loosened the screw. The fse tightened the arm and lubricated the joints which solved the reported issue. The fse repaired the impulse oscillatory spirometry (ios) arm and was returned to the customer. The device is currently working without issue.
Patient Sequence No: 1, Text Type: N, H10


[41419863] The customer reported the impulse oscillatory spirometry (ios) arm fell off. There was no patient involvement associated with this event. The carefusion field service engineer (fse) was scheduled to inspect the unit at the customer's facility.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2016-03378
MDR Report Key5527757
Date Received2016-03-25
Date of Report2016-02-29
Date of Event2016-02-29
Date Mfgr Received2016-02-29
Device Manufacturer Date2013-08-01
Date Added to Maude2016-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2016-03-25
Model NumberVMAX ENCORE 22
Catalog Number777404-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.