SURE STEP FOLEY TRAY SYSTEM A303416A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-29 for SURE STEP FOLEY TRAY SYSTEM A303416A manufactured by C.r. Bard, Inc..

Event Text Entries

[41313779]
Patient Sequence No: 1, Text Type: N, H10


[41313780] Rn reported bard stat lock to secure indwelling catheter was occluding the flow of urine into the urine meeter, as if it was clamped. The nurse reported 45 cc urine came out once the stat lock removed. She then used taped on the leg instead. The leg securement device called stat lock is part of the foley tray system. Manufacturer response for 16 fr sure step foley tray system, sure step foley tray system/ bard medical (per site reporter): the end user is not applying the tubing into the stat lock as instructed. There is only one way to put the tubing into the securement device and if the user puts tubing in the opposite direction this will crimp or occlude the outflow of urine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5531183
MDR Report Key5531183
Date Received2016-03-29
Date of Report2016-03-23
Date of Event2015-12-15
Report Date2016-03-23
Date Reported to FDA2016-03-23
Date Reported to Mfgr2016-03-23
Date Added to Maude2016-03-29
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSURE STEP FOLEY TRAY SYSTEM
Generic NameBAG, URINE, COLLECTION
Product CodeNNW
Date Received2016-03-29
Catalog NumberA303416A
Device AvailabilityN
Device Age0 DY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerC.R. BARD, INC.
Manufacturer Address8195 INDUSTRIAL BLVD. COVINGTON, GA 30014 US 30014


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-29