CHROMID(TM) CARBA SMART 414685

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-03-29 for CHROMID(TM) CARBA SMART 414685 manufactured by Biom?rieux Sa.

Event Text Entries

[41425516] A customer in (b)(6) reported false negative results when using; chromid carba smart. A patient rectal sample was run after admission to the hospital and no growth was observed after incubation. Three days after admission the blood culture of the patient returned positive at (b)(6), the results of the identified as; enterobacter cloacae (carbapenemase positive).
Patient Sequence No: 1, Text Type: D, B5


[47519746] An investigation into a false negative event while using the chromid? Carba smart test kit was performed. A complaint review for similar reports against on the lot and product number was performed. No systemic issue was identified as part of the trend analysis. Quality control strains were tested on 3 lots of test kits including the lot in question. This testing resulted in numeration, size of colonies and coloration intensity are within specification. The customer returned strain was tested on 5 separate lot of tests, varying over the shelf life of the product. This testing resulted in numeration, size of colonies and coloration intensity are within specification. Based on the results of the investigation, the complaint could not be confirmed and the product is operating within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00062
MDR Report Key5532592
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-03-29
Date of Report2016-03-12
Date Mfgr Received2016-03-12
Device Manufacturer Date2015-12-21
Date Added to Maude2016-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID(TM) CARBA SMART
Generic NameCHROMID(TM) CARBA SMART
Product CodeJSI
Date Received2016-03-29
Catalog Number414685
Lot Number1004537500
Device Expiration Date2016-03-06
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-29

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.