IMMUNOGLOBULIN E IGE 04827031190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,user facility report with the FDA on 2016-03-29 for IMMUNOGLOBULIN E IGE 04827031190 manufactured by Roche Diagnostics.

Event Text Entries

[41425415] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[41425416] The customer complained of erroneous results for 1 patient sample tested for immunoglobulin e (ige ii). It is not known if erroneous results were reported outside of the laboratory. The initial ige ii result was 925. 00 iu/ml. The sample was diluted x5 and the result was >12,500. 00 iu/ml. The sample was diluted x400 and the result was 163,623. 00 iu/ml. No adverse event occurred. The e602 analyzer serial number was not provided.
Patient Sequence No: 1, Text Type: D, B5


[45766373] The patient sample was submitted for investigation. During the investigation the customer's results were reproduced. The very high ige ii results causes a high dose hook effect as described in product labeling. Based on the investigation data, a general reagent issue can be excluded.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00369
MDR Report Key5532914
Report SourceFOREIGN,OTHER,USER FACILITY
Date Received2016-03-29
Date of Report2016-05-23
Date of Event2016-02-25
Date Mfgr Received2016-03-09
Date Added to Maude2016-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIMMUNOGLOBULIN E
Generic NameRADIOIMMUNOASSAY, IMMUNOGLOBULINS (D,E)
Product CodeJHR
Date Received2016-03-29
Model NumberIGE
Catalog Number04827031190
Lot Number181429
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-29

© 2026 FDA.report