OXYCON MOBILE 001-143250

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-03-30 for OXYCON MOBILE 001-143250 manufactured by Carefusion, Inc.

Event Text Entries

[41663501] This mdr is being submitted due to a retrospective complaint review performed by carefusion. The carefusion identification file is (b)(4). The device was evaluated at the customer's facility by a carefusion field service engineer (fse). He was unable to duplicate the reported event. As a correction, the fse installed the hotfix software and did a system calibration. Carefusion continues to track and trend any incident related to this issue. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[41663502] The customer reported the emergency stop function on their nordictrack x9i treadmill causes the (b)(4) software to shut off. When the software shuts off so does the ecg and breath by breath monitoring. There was no known patient involvement.
Patient Sequence No: 1, Text Type: D, B5


[43735235] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[43735236] The customer reported the emergency stop function on their nordictrack x9i treadmill causes the jlab software to shut off. When the software shuts off so does the ecg and breath by breath monitoring. The customer did not clarify if there was patient involvement or not, it is currently unknown.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2016-03409
MDR Report Key5536781
Date Received2016-03-30
Date of Report2014-03-31
Date of Event2014-03-31
Date Mfgr Received2016-04-21
Date Added to Maude2016-03-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone8473628056
Manufacturer G1CAREFUSION, INC
Manufacturer StreetLIEBNIZSTRASSE 7
Manufacturer CityHOECHBERG
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOXYCON MOBILE
Generic NameCALCULATOR, PULMONARY FUNCTION DATA
Product CodeBZC
Date Received2016-03-30
Model NumberOXYCON MOBILE
Catalog Number001-143250
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer AddressLIEBNIZSTRASSE 7 HOECHBERG 97204 GM 97204


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-30

© 2026 FDA.report