FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-03-31 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[41564336] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[41564337] The customer received a questionable free psa result from analytic e module serial number (b)(4). The initial total psa result was 6. 67 ng/ml and the free psa result was 10. 05 ng/ml. The repeat total psa result was 6. 73 ng/ml and the repeat free psa result was 2. 54 ng/ml. Information concerning if any erroneous result was reported outside the laboratory was requested, but it was unknown. The patient was not adversely affected.
Patient Sequence No: 1, Text Type: D, B5


[44458119] A specific root cause could not be identified. Sample from the patient was submitted for investigation. The issue could not be reproduced and no specific interference was found.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00390
MDR Report Key5538024
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-03-31
Date of Report2016-05-02
Date of Event2016-03-17
Date Mfgr Received2016-03-18
Date Added to Maude2016-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2016-03-31
Model NumberNA
Catalog Number03289788190
Lot Number188436
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-03-31