GALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE 1008112-0C NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2004-04-21 for GALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE 1008112-0C NA manufactured by Guidant, Houston.

Event Text Entries

[312240] The physicist reported the following scenario during a treatment: the site started to treat the first position in a two-position treatment. The system lost power during this treatment; therefore, the first position only received 1973 cgy dose. Instead of continuing this case, the oncologist repositioned the catheter to the second position and started a new two-position treatment. The second position received the normal 20 gy dose and then the system retracted to the third position and dwelled in this third position for a couple of seconds before the treatment. The second position received the normal 20 gy dose and then the system retracted to the third position and dwelled in this third position for a couple of seconds before the treatment was interrputed. This third area received a 56 cgy dose.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651662-2004-00004
MDR Report Key554040
Report Source05,06,07
Date Received2004-04-21
Date of Report2004-03-26
Date of Event2003-12-02
Date Mfgr Received2004-03-26
Device Manufacturer Date2000-07-01
Date Added to Maude2004-11-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHYSICIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDENISE VINMANS
Manufacturer Street26531 YNEZ RD
Manufacturer CityTEMECULA CA 92591
Manufacturer CountryUS
Manufacturer Postal92591
Manufacturer Phone9099142050
Manufacturer G1GUIDANT, HOUSTON
Manufacturer Street8934 KIRBY DR
Manufacturer CityHOUSTON TX 77054
Manufacturer CountryUS
Manufacturer Postal Code77054
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameGALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE
Generic NameRADIATION SYSTEM CARTRIDGE
Product CodeMOU
Date Received2004-04-21
Returned To Mfg2004-01-09
Model Number1008112-0C
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeNA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key543667
ManufacturerGUIDANT, HOUSTON
Manufacturer Address8934 KIRBY DR HOUSTON TX 77054 US
Baseline Brand NameGALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE
Baseline Generic NameRADIATION SYSTEM CARTRIDGE
Baseline Model No1008112-0C
Baseline Device FamilyRADIATION SYSTEM CARTRIDGE
Baseline Shelf Life ContainedN
Baseline Shelf Life [Months]NA
Baseline PMA FlagY
Premarket ApprovalP0000
Baseline 510K PMNN
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2004-04-21