CLINIMACS PLUS INSTRUMENT 151-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-04-01 for CLINIMACS PLUS INSTRUMENT 151-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[41665557] Device difficult to maintain - customer does not have preventive maintenance contract with miltenyi, the machine is very old, therefore difficult to maintain without routine maintenance. Process evaluation - miltenyi biotec service group has sent the customer a quote for service and is waiting for approval from customer. Service has quote out to customer.
Patient Sequence No: 1, Text Type: N, H10


[41665558] The customer received an error message #12 on their clinimacs plus instrument (b)(4)after the sample had been loaded. The operator was not next to the instrument when the error appeared, however, the negative fraction had already been eluted. No fluid/cells had been eluted into the positive fraction. Error 12 reads magnet error - refer to manual. Miltenyi technical service was called immediately and guided the customer through 2 emergency procedures. They were able to rescue cells, therefore no harm was caused to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2016-00061
MDR Report Key5540901
Report SourceHEALTH PROFESSIONAL
Date Received2016-04-01
Date of Report2016-04-01
Date of Event2016-01-27
Device Manufacturer Date2001-07-01
Date Added to Maude2016-04-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NANCY JOHANSEN
Manufacturer Street85 HAMILTON ST
Manufacturer CityCAMBRIDGE MA 02139
Manufacturer CountryUS
Manufacturer Postal02139
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS PLUS INSTRUMENT
Generic NameCLINIMACS PLUS INSTRUMENT
Product CodeOVG
Date Received2016-04-01
Catalog Number151-01
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STRASSE 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-01