MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2004-08-26 for BARD * 0603590 manufactured by Bard Acess Systems.
[21616038]
In 7/26/2004 pt had groshong/mri port insertion for breast cancer-chest x-ray showed tip in right atrium. Pt returned to surgery after one month because of port malfunction. Catheter was removed, and it was noted that there was a fracture of catheter at the level where the catheter passed over the clavicle with migration of the catheter.
Patient Sequence No: 1, Text Type: D, B5
Report Number | MW1033092 |
MDR Report Key | 554395 |
Date Received | 2004-08-26 |
Date of Report | 2004-08-25 |
Date of Event | 2004-08-16 |
Date Added to Maude | 2004-11-15 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | RISK MANAGER |
Health Professional | 3 |
Initial Report to FDA | 0 |
Report to FDA | 0 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | BARD |
Generic Name | GROSHONG CATHETER 8 FR. |
Product Code | DQD |
Date Received | 2004-08-26 |
Model Number | * |
Catalog Number | 0603590 |
Lot Number | 22E04518 |
ID Number | * |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 544023 |
Manufacturer | BARD ACESS SYSTEMS |
Manufacturer Address | 5245 W. AMELIA EARHART SALT LAKE CITY UT * US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2004-08-26 |