PROVOX VEGA 8133

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-07 for PROVOX VEGA 8133 manufactured by Atos Medical Ab.

Event Text Entries

[42888796] The esophagus flange is ruptured about 80% of the circumference, no parts are missing. The blue ring of the voice prosthesis shall be circular when seen from the esophagus side. The ring of the returned voice prosthesis is deformed; it is shaped as the letter "d". This deformation is most likely caused by a wrongly performed insertion/reload. Conclusion/action judged from the damage on the flange and blue ring this is not a result of a faulty product. The device failure is most likely caused by abnormal use during the insertion procedure. Corrective and preventive action: no product fault, corrective or preventive actions are not considered necessary except from informing the customer of the importance of following the instructions for use.
Patient Sequence No: 1, Text Type: N, H10


[42888797] During insertion procedure the clinician found that the esophageal flange of the voice prosthesis was torn and barely connecting with the shaft of the prosthesis. The event was discovered when the prosthesis came out of the loading tube at the end of the insertion. The prosthesis was removed right away.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8032044-2016-00002
MDR Report Key5554083
Date Received2016-04-07
Date of Report2016-04-07
Date of Event2016-02-28
Date Mfgr Received2016-03-03
Device Manufacturer Date2015-07-01
Date Added to Maude2016-04-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELIN ALGOTSSON
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer PostalSE-24222
Manufacturer Phone641519800
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer Postal CodeSE-24222
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePROVOX VEGA
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2016-04-07
Returned To Mfg2016-03-18
Catalog Number8133
Lot Number1511114
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressKRAFTGATAN 8 HORBY, SE-24222 SW SE-24222