HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT DFH-0008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-07 for HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT DFH-0008 manufactured by Microsurgical Technology Inc.

Event Text Entries

[42681086] The device evaluation showed that the scissor was used to cut to hard of an object, an iol. This scissor is indicated to cut soft tissue and objects in the eye per the dfu.
Patient Sequence No: 1, Text Type: N, H10


[42681087] Ophthalmic scissor blade broke in the eye while cutting a lens. The blade was removed without issue and there was no patient impact.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2016-00003
MDR Report Key5556336
Date Received2016-04-07
Date of Report2016-04-06
Date Mfgr Received2016-03-09
Device Manufacturer Date2012-02-17
Date Added to Maude2016-04-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT MAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT
Generic NameOPHTHALMIC SCISSOR
Product CodeHNF
Date Received2016-04-07
Returned To Mfg2016-03-24
Model NumberDFH-0008
Catalog NumberDFH-0008
Lot Number045283
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.