SURE LOC XPS ARM POSITIONER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-08 for SURE LOC XPS ARM POSITIONER manufactured by Steris Corporation - Montgomery.

Event Text Entries

[42872506] The facility stated that on multiple occasions, the unit was positioned, and minutes later, the positioner dropped approximately once inch. A steris service technician arrived at the facility, inspected the unit, and identified the trigger pin did not seem fully engaged in the unit's locking mechanism. This would cause the unit to slip to the next locking position as reported by the facility. The unit was removed from service and returned to the 3rd party manufacturer for further evaluation. A replacement unit has since been provided to the user facility; no further issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[42872507] The facility reported that the sure loc arm positioner did not operate to specification during a patient procedure. No injury, procedural delay, or cancellation was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1043572-2016-00036
MDR Report Key5563998
Date Received2016-04-08
Date of Report2016-04-08
Date of Event2016-03-14
Date Mfgr Received2016-03-14
Date Added to Maude2016-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSURE LOC XPS ARM POSITIONER
Generic NameTABLE ACCESSORY
Product CodeFSE
Date Received2016-04-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.