DIMENSION? CLINICAL CHEMISTRY SYSTEMS RF420, SMN10444904

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-04-08 for DIMENSION? CLINICAL CHEMISTRY SYSTEMS RF420, SMN10444904 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[42224352] The cause of the falsely elevated mmb results is unknown. After the customer observed elevated results on four patient samples and found that level 1 qc was out, customer migrated to a new reagent cartridge, reran qc and confirmed it was within range, and repeated the samples. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[42224353] Discordant elevated mass creatine kinase (mmb) results were obtained on four patient samples on the dimension exl system and results were reported. The laboratory decided to check qc and level 1 qc was high. The samples were repeated on a new reagent cartridge and lower results were obtained. Corrected reports were sent. Patient treatment was not altered or prescribed on the basis of the discordant elevated mass creatine kinase (mmb) results. There was no report of adverse health consequences as a result of the discordant elevated mass creatine kinase (mmb) results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2016-00227
MDR Report Key5564325
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-04-08
Date of Report2016-03-30
Date of Event2016-03-30
Date Mfgr Received2016-03-30
Device Manufacturer Date2015-10-05
Date Added to Maude2016-04-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEDWARD SZYMANSKI
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026317672
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE
Manufacturer CityNEWARK DE 19714
Manufacturer CountryUS
Manufacturer Postal Code19714
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION? CLINICAL CHEMISTRY SYSTEMS
Generic NameMASS CREATINE KINASE MB ISOENZYME FLEX? REAGENT CARTRIDGE
Product CodeJHY
Date Received2016-04-08
Catalog NumberRF420, SMN10444904
Lot NumberFB6279
Device Expiration Date2016-10-05
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE NEWARK DE 19714 US 19714


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.