[42614071] While performing forceps biopsy a noise was heard, instrument removed and inspected, revealing broken biting mechanism. Wires holding jaws together (normal on bottom) broke, allowing the jaws to pivot (top forceps), increasing the size approximately 3-4x and exposing sharp edges.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5573049
MDR Report Key5573049
Date Received2016-04-11
Date of Report2016-03-23
Date of Event2016-03-02
Date Facility Aware2016-03-02
Report Date2016-03-29
Date Reported to FDA2016-04-01
Date Reported to Mfgr2016-04-01
Date Added to Maude2016-04-13
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBOSTON SCIENTIFIC, RADIAL JAW 4
Generic NameBIOPSY FORCEPS
Product CodeBWH
Date Received2016-04-11
Returned To Mfg2016-03-15
Model NumberRADIAL JAW 4
Catalog NumberM00515180
Lot Number18546277
Device Expiration Date2018-10-27
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC
Manufacturer Address300 BOSTON SCIENTIFIC WAY MALBOROUGH MA 017521234 US 017521234


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2016-04-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.