MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-18 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[42953785] This event occurred in (b)(6). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[42953786] The customer reported that they received erroneous results for one patient sample tested for total prostate-specific antigen (tpsa) and free prostate-specific antigen (fpsa) on an e601 analyzer. The fpsa result was greater than the tpsa result. This medwatch will cover fpsa. Please refer to the medwatch with patient identifier (b)(6) for information related to tpsa. The sample resulted as 0. 149 ng/ml for tpsa and 2. 59 ng/ml for fpsa. The results were reported outside of the laboratory to the patient. The patient was not adversely affected. The e601 analyzer serial number was asked for, but not provided.
Patient Sequence No: 1, Text Type: D, B5


[44897592] A sample from the patient was provided for investigation. Investigations were able to confirm the customer's results. Investigations performed with an antibody that has a different specificity than the one used in the tpsa assay increased tpsa recovery of the sample significantly. Further investigations could not be performed since there was not enough sample remaining.
Patient Sequence No: 1, Text Type: N, H10


[47225049] A new sample from the same patient has been provided for investigation. This sample also had erroneous results for fpsa and tpsa. This sample had a fpsa result of 3. 180 ng/ml and a tpsa result of 0. 220 ng/ml. The erroneous results were not reported outside of the laboratory. The patient was not adversely affected.
Patient Sequence No: 1, Text Type: N, H10


[48207300] Investigations of the new patient sample were able to confirm the customer's results. For investigations, a total psa reagent with different analyte specificity was used. Upon use of this reagent, a significant increase in total psa recovery of the sample can be seen. The result of total psa with this reagent is higher than the free psa result. The observed behavior may be explained by the presence of auto-antibodies blocking the epitope on complexed psa or an anti-idiotype antibody which blocks the catcher antibody of the total psa assay. Product labeling indicates that psa isoforms do exist which may be measured differently by different psa tests.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00446
MDR Report Key5584535
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-04-18
Date of Report2016-06-22
Date of Event2016-03-30
Date Mfgr Received2016-03-31
Date Added to Maude2016-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2016-04-18
Model NumberNA
Catalog Number03289788190
Lot Number188436
ID NumberNA
Device Expiration Date2017-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.