CHROMID? CPS ELITE AGAR PLATES 416172

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-19 for CHROMID? CPS ELITE AGAR PLATES 416172 manufactured by Biom?rieux Sa.

Event Text Entries

[43018701] A customer in (b)(6) notified biomerieux of a discrepant false negative result associated with chromid cps elite agar plates (reference (b)(4)the customer reported the media shrinking and results with no growth. Incubation was confirmed at 37 deg c for 24 hours. The customer indicated patient results were not affected. In addition, the customer indicated there was no evidence to support erroneous results were reported to a physician, patients were treated incorrectly or harmed, and/or delays in reporting results.
Patient Sequence No: 1, Text Type: D, B5


[46504921] A customer in (b)(6) notified biom? Rieux of a discrepant false negative result associated with chromid? Cps elite agar plates (reference (b)(4)). The customer reported the media shrinking and results with no growth. An internal biom? Rieux investigation was performed with results as follows: no other complaints linked to batch 1004658500. Problem determined to not be related to customer error/incorrect incubation. Product identified as being more sensitive to shrinkage than other culture media. Chromid? Cps elite shrinkage issue associated with capa (b)(4). Internal corrective actions implemented as a result of the investigation conclusions: position chromid? Cps elite agar plates from lid down to lid up position to unify the way customers. Are receiving plates and help decrease retraction issues.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00068
MDR Report Key5587237
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-04-19
Date of Report2016-04-15
Date Mfgr Received2016-04-15
Date Added to Maude2016-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? CPS ELITE AGAR PLATES
Generic NameCHROMID? CPS ELITE AGAR PLATES
Product CodeJXA
Date Received2016-04-19
Catalog Number416172
Lot Number1004658500
Device Expiration Date2016-06-12
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.