OLYMPUS VISERA ELITE XENON LIGHT SOURCE CLV-S190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,literature,user facil report with the FDA on 2016-04-20 for OLYMPUS VISERA ELITE XENON LIGHT SOURCE CLV-S190 manufactured by Olympus Medical Systems Corp..

Event Text Entries

[43167428] The subjected device was returned to olympus medical systems corp. (omsc) for evaluation. Omsc evaluated the device and found that the phenomenon was reproduced. Furthermore omsc found that the power-supply unit did not output the lighting voltage for the xenon lamp. Omsc evaluated the power-supply unit and found that the resistance mounted to the circuit board had the marks of heat and the resistance was broken. Omsc replaced this resistance with another one. The power-supply unit outputted the lighting voltage for the xenon lamp appropriately. So omsc surmised that the breakage of this resistance was the cause of this phenomenon. Omsc surmised that the cause of the breakage of the resistance is the following in theory. Any foreign material (e. G. Dust, smother) was attached to a high-voltage generating circuit of the unit that is supplied a voltage from this power-supply unit. An electric charge leaked through any foreign material. The power-supply unit supplied a large voltage unintentionally. As a result, an electrical load in the circuit of the power-supply unit had intensified. The resistance developed fever and the resistance was broken. The (b)(4) instruction manual states the corresponding method when the xenon lamp went out and the cleaning procedure for the device. There were no further details provided. If significant additional information is received, this report will be supplemented. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


[43167429] Olympus was informed that the xenon lamp went out and the emergency lamp lighted up during an unspecified ent procedure. As a result of investigation, olympus was informed the following information on march 25th 2016. At this case, the facility replaced the (b)(4) with a spare device made by (b)(4), and completed the procedure. There was no report of the patient's injury regarding this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2016-00494
MDR Report Key5589791
Report SourceFOREIGN,LITERATURE,USER FACIL
Date Received2016-04-20
Date of Report2016-04-20
Date of Event2015-11-25
Date Mfgr Received2016-03-25
Device Manufacturer Date2015-02-09
Date Added to Maude2016-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. YUTAKA YANAGAWA
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 1928507
Manufacturer CountryJA
Manufacturer Postal1928507
Manufacturer Phone42 6425177
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VISERA ELITE XENON LIGHT SOURCE
Generic NameLIGHT SOURCE
Product CodeGCT
Date Received2016-04-20
Returned To Mfg2016-03-07
Catalog NumberCLV-S190
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-04-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.