ROHO? STANDARD CUSHION COVER COV-MTO

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2016-04-21 for ROHO? STANDARD CUSHION COVER COV-MTO manufactured by Roho, Inc..

Event Text Entries

[43207060] The suspect product was not available for evaluation. During conversations with the complainant, the following was noted: the customer does not follow the cushion adjustment instructions in the operation manual (see attached cushion adjustment instructions from the cushion and cover operation manual, revision 4/21/2015). Instead, while he is off the cushion, he pushes both fists into the cushion and as long as he doesn't touch the base, he uses the cushion. This method could result in the cushion being over-inflated and effect his stability while on the cushion and cover. Based on communications with the complainant, roho has determined that the cushion cover did not malfunction in any way. Production logs for the cover were reviewed and show that it was made per the documented requirements. See the attached productions log review file. During conversations with the complainant, roho representatives attempted to obtain the more precise manufacturer's information associated with the manual wheelchair in question. The customer was unable to provide this information.
Patient Sequence No: 1, Text Type: N, H10


[43207061] The customer stated he was going slowly down a sidewalk and hit a curb cut (where the curb meets the street) and slid out of the chair. He stated that he usually does a wheelie to get down or up curb cuts, but this time he was talking with a friend and didn't notice the curb in time to do the wheelie. Also, he stated the curb cut was a little steeper than a normal curb cut. After moving over the curb cut, the customer stated he was unable to maintain his position. He stopped his chair, but he kept sliding. It was a slow slide off of the chair. He also stated that he slipped twice that day, but got caught by someone one of the times. The customer stated he got cut up and rolled to the left and out of the chair. When he hit the ground, he fractured his tibia, broke off a small part of his patella and damaged ligament. He was seen by his physician, who stated surgery was not needed. He needs to keep his leg up and ice it a few times a day for several weeks. He also switched to a high protein diet. He is strapping his legs in now because he is at high risk of an additional injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419507-2016-00001
MDR Report Key5594127
Report SourceCONSUMER
Date Received2016-04-21
Date of Report2016-04-15
Date of Event2016-01-02
Date Mfgr Received2016-03-24
Date Added to Maude2016-04-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ROBIN GERGEN
Manufacturer Street1501 S. 74TH ST.
Manufacturer CityBELLEVILLE IL 62223
Manufacturer CountryUS
Manufacturer Postal62223
Manufacturer Phone8003562990
Manufacturer G1ROHO, INC.
Manufacturer Street100 N. FLORIDA AVE.
Manufacturer CityBELLEVILLE IL 62221
Manufacturer CountryUS
Manufacturer Postal Code62221
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROHO? STANDARD CUSHION COVER
Generic NameACCESSORIES, WHEELCHAIR
Product CodeKNO
Date Received2016-04-21
Model NumberCOV-MTO
Lot Number20151209
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROHO, INC.
Manufacturer Address100 N. FLORIDA AVE. BELLEVILLE IL 62221 US 62221


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-04-21

© 2026 FDA.report