SUPPORT ARM 177 6481720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-22 for SUPPORT ARM 177 6481720 manufactured by Maquet Critical Care Ab.

Event Text Entries

[44190713] It was reported that the support arm broke at one of the joints. There was no patient harm reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[47938759] Our investigation into this complaint has now been finalized. It was reported that a support arm for a ventilator broke at one of the joints. Received pictures of the broken support arm confirmed the event description. Our evaluation of the received pictures showed that the support arm broke at the joint nearest to the bracket. The support arm is a casting, and it most probably developed a crack at an earlier occasion either by overloading or impact, and this led to the reported breakage. Previous investigations of similar faults led to a redesign and change of the manufacturing process in order to obtain a higher mechanical strength of the support arm. The date stamp showed that the support arm associated with this complaint had been manufactured before this change was implemented in production.
Patient Sequence No: 1, Text Type: N, H10


[47938760] (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010042-2016-00224
MDR Report Key5597818
Date Received2016-04-22
Date of Report2016-02-19
Date of Event2016-02-08
Date Mfgr Received2016-02-19
Device Manufacturer Date2009-09-01
Date Added to Maude2016-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Manufacturer G1MAGNUS LINDQVIST
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPPORT ARM 177
Generic NameSUPPORT, ARM
Product CodeIOY
Date Received2016-04-22
Catalog Number6481720
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.