ACCUTREND CHOLESTEROL 05213312160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-04-26 for ACCUTREND CHOLESTEROL 05213312160 manufactured by Roche Diagnostics.

Event Text Entries

[43609665] .
Patient Sequence No: 1, Text Type: N, H10


[43609666] The reporter stated that a cholesterol value of 238 mg/dl was obtained on the accutrend meter (serial number (b)(4)). One week later a laboratory draw was done and the result was 146 mg/dl. The laboratory draw was done as a follow-up to the meter result. There were no treatments given or delayed. The controls on the meter were run and they were in range before and after the stated result. It was stated that the patient had normal physical activity and food and drink. The patient is doing well. There was no adverse event. The suspect product was requested to be returned and the replacement product was sent.
Patient Sequence No: 1, Text Type: D, B5


[46696482] A specific root cause for the event could not be determined. The product was not returned for investigation. Retention samples were tested with control solution. The results were within the range and there were no abnormalities. The customer product was not returned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00489
MDR Report Key5607406
Report SourceHEALTH PROFESSIONAL
Date Received2016-04-26
Date of Report2016-06-06
Date of Event2016-03-08
Date Mfgr Received2016-03-31
Date Added to Maude2016-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCUTREND CHOLESTEROL
Generic NameCHOLESTEROL (TOTAL) TEST SYSTEM
Product CodeCHH
Date Received2016-04-26
Model NumberNA
Catalog Number05213312160
Lot Number20402521
ID NumberNA
Device Expiration Date2016-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.