MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-27 for COBAS E-ELECSYS ANTI-CCP 05031656190 manufactured by Roche Diagnostics.
        [43854729]
This event occurred in (b)(6).
 Patient Sequence No: 1, Text Type: N, H10
        [43854730]
The customer complained of erroneous results for 1 patient tested for anti-ccp. The erroneous results were reported outside of the laboratory. The initial anti-ccp result from the e411 analyzer was 30. 79 u/ml. This result was reported outside of the laboratory. A new sample was obtained 3 days later and run on the e411 analyzer again and the result was "the same. " the initial sample was repeated on a different e411 analyzer at a different customer site and the result was "the same. " the initial sample was repeated on an abbott architect instrument and the result was 1. 5 u/ml. A new sample was obtained and the result from a siemens system was <1. 2 u/ml. No adverse event occurred. The e411 analyzer serial number was not provided.
 Patient Sequence No: 1, Text Type: D, B5
        [45295234]
A specific root cause could not be identified. Additional information for further investigation was requested but was not provided.
 Patient Sequence No: 1, Text Type: N, H10
| Report Number | 1823260-2016-00523 | 
| MDR Report Key | 5612405 | 
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U | 
| Date Received | 2016-04-27 | 
| Date of Report | 2016-05-11 | 
| Date of Event | 2016-04-01 | 
| Date Mfgr Received | 2016-04-07 | 
| Date Added to Maude | 2016-04-27 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | NA MICHAEL LESLIE | 
| Manufacturer Street | 9115 HAGUE ROAD NA | 
| Manufacturer City | INDIANAPOLIS IN 46250 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 46250 | 
| Manufacturer Phone | 3175214343 | 
| Manufacturer G1 | ROCHE DIAGNOSTICS GMBH | 
| Manufacturer Street | SANDHOFERSTRASSE 116 NA | 
| Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 | 
| Manufacturer Country | GM | 
| Manufacturer Postal Code | 68305 | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Removal Correction Number | NA | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | COBAS E-ELECSYS ANTI-CCP | 
| Generic Name | ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) | 
| Product Code | NHX | 
| Date Received | 2016-04-27 | 
| Model Number | NA | 
| Catalog Number | 05031656190 | 
| Lot Number | 188366 | 
| ID Number | NA | 
| Device Expiration Date | 2016-05-31 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | Y | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | R | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | ROCHE DIAGNOSTICS | 
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2016-04-27 |