COBAS E-ELECSYS ANTI-CCP 05031656190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-27 for COBAS E-ELECSYS ANTI-CCP 05031656190 manufactured by Roche Diagnostics.

Event Text Entries

[43854729] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[43854730] The customer complained of erroneous results for 1 patient tested for anti-ccp. The erroneous results were reported outside of the laboratory. The initial anti-ccp result from the e411 analyzer was 30. 79 u/ml. This result was reported outside of the laboratory. A new sample was obtained 3 days later and run on the e411 analyzer again and the result was "the same. " the initial sample was repeated on a different e411 analyzer at a different customer site and the result was "the same. " the initial sample was repeated on an abbott architect instrument and the result was 1. 5 u/ml. A new sample was obtained and the result from a siemens system was <1. 2 u/ml. No adverse event occurred. The e411 analyzer serial number was not provided.
Patient Sequence No: 1, Text Type: D, B5


[45295234] A specific root cause could not be identified. Additional information for further investigation was requested but was not provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-00523
MDR Report Key5612405
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-04-27
Date of Report2016-05-11
Date of Event2016-04-01
Date Mfgr Received2016-04-07
Date Added to Maude2016-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS E-ELECSYS ANTI-CCP
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2016-04-27
Model NumberNA
Catalog Number05031656190
Lot Number188366
ID NumberNA
Device Expiration Date2016-05-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.