MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-27 for COBAS E-ELECSYS ANTI-CCP 05031656190 manufactured by Roche Diagnostics.
[43854729]
This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[43854730]
The customer complained of erroneous results for 1 patient tested for anti-ccp. The erroneous results were reported outside of the laboratory. The initial anti-ccp result from the e411 analyzer was 30. 79 u/ml. This result was reported outside of the laboratory. A new sample was obtained 3 days later and run on the e411 analyzer again and the result was "the same. " the initial sample was repeated on a different e411 analyzer at a different customer site and the result was "the same. " the initial sample was repeated on an abbott architect instrument and the result was 1. 5 u/ml. A new sample was obtained and the result from a siemens system was <1. 2 u/ml. No adverse event occurred. The e411 analyzer serial number was not provided.
Patient Sequence No: 1, Text Type: D, B5
[45295234]
A specific root cause could not be identified. Additional information for further investigation was requested but was not provided.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1823260-2016-00523 |
MDR Report Key | 5612405 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2016-04-27 |
Date of Report | 2016-05-11 |
Date of Event | 2016-04-01 |
Date Mfgr Received | 2016-04-07 |
Date Added to Maude | 2016-04-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS E-ELECSYS ANTI-CCP |
Generic Name | ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) |
Product Code | NHX |
Date Received | 2016-04-27 |
Model Number | NA |
Catalog Number | 05031656190 |
Lot Number | 188366 |
ID Number | NA |
Device Expiration Date | 2016-05-31 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-04-27 |