TISHLER MORGAN BIOPSY PUNCH 64-689

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-27 for TISHLER MORGAN BIOPSY PUNCH 64-689 manufactured by Coopersurgical, Inc..

Event Text Entries

[57863833] Coopersurgical inc. Is currently investigating the reported complaint condition. The device involved in the complaint has not been returned by the customer for evaluation. Once the investigation is completed a follow-up report will be filed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[57863834] "punches dull/doctor unable to get good sample. " "patient had significant pain and bleeding trying to get a biopsy. " (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[58370781] On 07/15/2015 coopersurgical inc. Is currently investigating the reported complaint condition. The device involved in the complaint has not been returned by the customer for evaluation. Once the investigation is completed a follow-up report will be filed. (b)(4). On 10/22/2015 investigation summary report: review dhr: inspect returned samples. The returned sample was a warranty repair item. The returned instrument was reviewed by a service and repair technician and found the punch needed to be sharpened (see repair order (b)(4)). This part is purchased from a supplier and is packaged at coopersurgical. This particular instrument was packaged in nov. 2014 and sent to the customer in dec. 2014. A review of the dhr (b)(4) did not find any abnormalities. A review of two year complaint history found two similar complaints. The punch is a known wear item from use and requires sharpening.
Patient Sequence No: 1, Text Type: N, H10


[58370782] "punches dull/doctor unable to get good sample. " "patient had significant pain and bleeding trying to get a biopsy. " (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1216677-2015-00056
MDR Report Key5612461
Date Received2016-04-27
Date of Report2015-08-10
Date of Event2015-07-15
Date Mfgr Received2015-07-21
Device Manufacturer Date2014-11-01
Date Added to Maude2016-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. NANA BANAFO
Manufacturer Street75 CORPORATE DRIVE
Manufacturer CityTRUMBULL CT 06611
Manufacturer CountryUS
Manufacturer Postal06611
Manufacturer Phone2036015200
Manufacturer G1COOPERSURGICAL, INC.
Manufacturer Street75 CORPORATE DRIVE
Manufacturer CityTRUMBULL CT 06611
Manufacturer CountryUS
Manufacturer Postal Code06611
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTISHLER MORGAN BIOPSY PUNCH
Generic NameTISHLER MORGAN BIOPSY PUNCH
Product CodeHFB
Date Received2016-04-27
Returned To Mfg2015-07-15
Model Number64-689
Catalog Number64-689
Lot Number4175
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOPERSURGICAL, INC.
Manufacturer Address75 CORPORATE DRIVE TRUMBULL CT 06611 US 06611


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-27