STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-04-28 for STRYKERVISION FLAT-PANEL MOUNTING ARM 0240095200 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[43888559] (b)(4). The device manufacture date is not known at this time. However, should it become available it will be provided in future reports. Additional information will be provided once the investigation has been completed
Patient Sequence No: 1, Text Type: N, H10


[43888560] It was reported the mintor fell off of the tower and hit a nurse in the head.
Patient Sequence No: 1, Text Type: D, B5


[51354395] (b)(4). Since device was not returned and lot number was not provided manufacture date is not known. Alleged failure: monitor fell off a tower. Probable root cause: manufacturing nonconformity (tekna); nonconforming vesa mounting plate; weld fracture; damage during shipping /handling; missing or damaged fasteners/screws; use error. The product was not returned for investigation therefore the reported failure mode was not confirmed. The failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10


[51354396] It was reported the mintor fell off of the tower and hit a nurse in the head.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2016-00407
MDR Report Key5615064
Report SourceUSER FACILITY
Date Received2016-04-28
Date of Report2016-04-05
Date of Event2016-04-05
Date Mfgr Received2016-04-05
Date Added to Maude2016-04-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KIMBERLY LYNCH
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRYKERVISION FLAT-PANEL MOUNTING ARM
Generic NameCART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Product CodeBZN
Date Received2016-04-28
Catalog Number0240095200
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-04-28

© 2026 FDA.report