THINPREP 5000 PROCESSOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-04-28 for THINPREP 5000 PROCESSOR manufactured by Hologic, Inc.

Event Text Entries

[45097910] A customer in (b)(6) reported their thinprep 5000 processor etched a wrong vial id onto 2 slides. Instrument operational, (b)(4) field service engineer (fse) dispatched. The instrument processed normally without any error presented to the operator. Fse noted 1d label on the vial, 2d label on the slide. The customer confirmed no patients needed to be recalled as a result of this incident as they were able to reprocess the slide with the correct label printed on the slide. Fse performed preventative maintenance per technical documentation. Instrument operational. Although no patients needed to be recalled as a result of this incident this is a reportable event since the instrument did not perform as intended.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2016-00105
MDR Report Key5615762
Date Received2016-04-28
Date of Report2016-04-28
Date Mfgr Received2016-03-29
Date Added to Maude2016-04-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSIDRA PIRACHA
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5082638884
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTHINPREP 5000 PROCESSOR
Generic NameTHINPREP 5000 PROCESSOR
Product CodeMKQ
Date Received2016-04-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC
Manufacturer Address250 CAMPUS DRIVE MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-04-28

© 2026 FDA.report