ETEST? METRONIDAZOLE MZH 530018

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-04-28 for ETEST? METRONIDAZOLE MZH 530018 manufactured by Biomerieux, Sa.

Event Text Entries

[44003359] A customer in sweden notified biom? Rieux of a discrepant results associated with etest? Metronidazole mzh 256 (reference 530018). The customer reported always getting low minimal inhibitory concentration (mic) when doing controls. The customer identified the control strain as e. Coli atcc 25285. The approved spreading interval was 0. 25-1 mg/l and an anaerobe plate without antibiotics was utilized. An investigation into the event will be initiated by biomerieux.
Patient Sequence No: 1, Text Type: D, B5


[51263432] An investigation into a discrepant result event while using the etest? Metronidazole was performed. The customer did not return their strain for testing at biomerieux. Quality control samples with atcc? 25922, atcc? 29741 and atcc? 25285 were run with retain samples from five (5) lots etest? Metronidazole test kits. The results of all testing were within specifications for all samples. Based on the results of the investigation, the complaint could not be confirmed and the product is operating within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00076
MDR Report Key5616965
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-04-28
Date of Report2016-04-05
Date Mfgr Received2016-04-05
Date Added to Maude2016-04-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE, 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? METRONIDAZOLE MZH
Generic NameETEST? METRONIDAZOLE MZH
Product CodeJWY
Date Received2016-04-28
Catalog Number530018
Lot Number1003409180
Device Expiration Date2019-09-07
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE, 38390 FR 38390


Patients

Patient NumberTreatmentOutcomeDate
10 2016-04-28

© 2026 FDA.report