CHROMID CARBA SMART AGAR 414685

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-05-02 for CHROMID CARBA SMART AGAR 414685 manufactured by Biom?rieux Sa.

Event Text Entries

[44167308] A customer in (b)(6) notified biomerieux of a discrepant result associated with chromid carba smart agar (reference 414685). The plate was inoculated with a quality control sample of e. Coli indicated as not being atcc strain, but from a microbiologist's carbapenemase collection that was isolated in tunisia in 2012 and produces tem, ctx-m-14 and oxa-244 enzymes. The customer indicated the organism is subbed from frozen each time qc is set up. The customer reported the organism grew scant pink on the carba side and there was no growth on the oxa side whereas, there should have only been growth on the oxa side.
Patient Sequence No: 1, Text Type: D, B5


[46119113] A customer in (b)(6) notified biom? Rieux of a discrepant result associated with chromid? Carba smart agar (reference 414685) involving a quality control sample of e. Coli where the organism grew scant pink on the carba side and there was no growth on the oxa side. An internal biom? Rieux investigation was performed with results as follows: complaint history review revealed no additional complaints involving the customer's lot. Biom? Rieux requested the strain from the customer; however, the customer's strain was not received. A retention sample of the customer's lot (1004671720) in parallel with two lots of chromid? Carba smart agar (1004737760/1004760850) and a lot of cos were tested. Testing strains included escherichia coli2 (ref (b)(4)), proteus rettgeri2 (ref (b)(4)), klebsiella pneumoniae2 (ref (b)(4)) nctc (b)(4), enterobacter cloacae (ref (b)(4)), klebsiella pneumoniae2 (ref (b)(4)), klebsiella pneumoniae2 (ref (b)(4)), and klebsiella pneumoniae2 (ref (b)(4)). Results- after incubation under aerobic conditions at 33-37? C for 18-24 hours, product performance was within specifications regarding fertility and color intensity. There were no differences between lots detected. All positive strains were recovered and no false negative reactions were detected.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00078
MDR Report Key5623945
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-05-02
Date of Report2016-04-15
Date Mfgr Received2016-04-15
Device Manufacturer Date2016-02-19
Date Added to Maude2016-05-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID CARBA SMART AGAR
Generic NameCHROMID CARBA SMART AGAR
Product CodeJSI
Date Received2016-05-02
Catalog Number414685
Lot Number1004671720
Device Expiration Date2016-05-05
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-02