CLINIMACS CD34 REAGENT N/A 171-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-05-05 for CLINIMACS CD34 REAGENT N/A 171-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[44677094] Miltenyi biotec (b)(4) has handed over an investigation questionnaire to the customer to obtain more details about the complained clinimacs? Cell separation. Based on the entire data and the facs plots of the positive fraction received, following results can be ascertained: the source product for the selection procedure was a fresh, mobilized leukapheresis product containing 27xl0e9 tnc and 76,8xl0e6 cd34-cells (= 0,28% cd34+). The separation was performed with one vial of clinimacs? Cd34 reagent, clinimacs? Tubing set and cd34 selection! -program. Igg was added prior labeling for blocking of unspecific binding sites. Small aggregates were observed in the original fraction positive fraction: viability was 94% cd34+ recovery was indeed 31 % purity was 94% results #: the selection procedure has been performed inside the specifications of the clinimacs? Cd34 reagent and the respective tubing set. The plt-wash was not exactly done at a 1:3 dilution. By using a lower amount of buffer, the platelet removal might not have been sufficient with potential consequence of weak labeling and loss of cd34+-cells.
Patient Sequence No: 1, Text Type: N, H10


[44677095] The customer complained that the recovery of cd34+-cells was low (3. 3%) for a clinimacs separation procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2016-00065
MDR Report Key5634846
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-05-05
Date of Report2016-05-04
Date of Event2016-03-11
Date Mfgr Received2016-04-06
Device Manufacturer Date2015-12-21
Date Added to Maude2016-05-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NANCY JOHANSEN
Manufacturer Street85 HAMILTON ST
Manufacturer CityCAMBRIDGE MA 02139
Manufacturer CountryUS
Manufacturer Postal02139
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS CD34 REAGENT
Generic NameCLINIMACS CD34 REAGENT
Product CodeOVG
Date Received2016-05-05
Model NumberN/A
Catalog Number171-01
Lot Number6151218004
Device Expiration Date2016-05-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STRASSE 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-05