ETEST? DAPTOMYCIN DPC 256 WW S30 412324

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-05-06 for ETEST? DAPTOMYCIN DPC 256 WW S30 412324 manufactured by Biomerieux, Sa.

Event Text Entries

[44595490] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[44595491] A customer in (b)(6) reported a daptomycin qc (atcc 29212 tm) failure oorl (out of range low) for enterococcus faecalis when tested with the etest daptomycin kit. Expected mic value is 1 - 4 g/ml; results obtained were 0. 5 g/ml and 0. 75 g/ml. There is no indication or report from the hospital to biomerieux that the daptomycin qc failure led to any adverse event related to a patient's state of health. There is no patient directly associated with the qc sample. Culture submittals have been requested by biomerieux for internal investigation. An internal biomerieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


[54646667] This report was initially submitted following notification that a customer in (b)(6)reported occurrence of a low mic result for enterococcus faecalis atcc? 29212? In association with the etest? Daptomycin kit. Biom? Rieux investigation was conducted. Expected values: 1-4 g/ml. Customer results: between (b)(6) 2016, four (4) results at 0. 5 g/ml on twelve (12) data. Four (4) data at 0. 75 g/ml are acceptable - instruction for use 9302553c indicates: an etest? Mic value which falls between standard two-fold dilutions must be rounded up to the next upper two-fold value before categorization. The customer did not submit test strips for investigational testing. Retained samples (customer batches 1004340160 and 1004093520) were tested, in parallel with another batch as reference. The investigation did not reproduce the issue observed by the customer with the quality control strains. The conformity of this batch was confirmed with three (3) qc atcc? Strains: atcc? 29212? Enterococcus feacalis, atcc? 29213? Staphylococcus aureus and atcc? 49619? Streptococcus pneumoniae. The investigation concluded the etest? Daptomycin kit is performing as intended. As the customer product was not submitted for internal biom? Rieux testing, it can be hypothesized: media: - as explained in cis 014 or 9302553c, daptomycin requires physiologic levels of calcium (ca2+) for expression of adequate activity. The bioavailable free ca2+ levels in brands and batches of mha powder and commercial mha plates can vary significantly and affect etest? Dpc results. - use only mha plates with consistent and appropriate ca2+ levels (25-40 g/ml). - perform quality control (qc) using both enterococcus faecalis atcc? 29212? (1-4) and staphylococcus aureus atcc? 29213? (0. 25-1 [ g/ml). Use of two qc strains allows verification of results within the susceptible range and around the breakpoint. When results are out of qc specifications or consistently skewed at the limits, repeat the test using a different brand of mha. Reading: the investigation observed microcolonies with enterococcus feacalis atcc? 29212?. => read different etest? Mic endpoint patterns using the etest? Reading guidelines = cis014. In this case, read where the microcolonies are inhibited.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2016-00087
MDR Report Key5636742
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-05-06
Date of Report2016-04-27
Date Mfgr Received2016-04-27
Device Manufacturer Date2015-10-02
Date Added to Maude2016-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE, 38390
Manufacturer CountryFR
Manufacturer Postal Code38390
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? DAPTOMYCIN DPC 256 WW S30
Generic NameETEST? DAPTOMYCIN
Product CodeJWY
Date Received2016-05-06
Catalog Number412324
Lot Number1004340160
Device Expiration Date2018-10-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE, 38390 FR 38390