[44753254] The dental office comingles their brushes, so we were unable to determine the lot or expiration date of the brush involved in this complaint. The dental office did send in a handful of brushes from the drawer where they comingle (lot and expiration also unable to be determined) for evaluation. During evaluation, we were unable to replicate the actual complaint when run on a handpiece per manufacturer's instructions. Some bristles were only able to be removed using considerable force and pliers. Due to a lack of information it can only be hypothesized that this event occurred due to the age of the brush, possibly an expired lot where the glue became brittle, or that an isolated event occurred where the brush in the complaint did not receive enough glue. No other complaints of this type have been received.
Patient Sequence No: 1, Text Type: N, H10


[44753255] "sent from my iphone per (b)(6): patient (she) (b)(6)- used brush to get stain off- once in mouth bristles went everywhere - including eye- which patient went to eye dr. And dr. Said ok - dentist gave patient a credit- but says everything is fine- she has been in this bus 33 yrs never seen it before sending a few back in the mail does not have lot # and is in no way upset. I will send out replacement do not know since they mix them that had expired"
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1941138-2016-00013
MDR Report Key5641376
Report SourceUSER FACILITY
Date Received2016-05-09
Date of Report2016-04-15
Date Mfgr Received2016-04-13
Date Added to Maude2016-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SARAH MOORE
Manufacturer Street13705 SHORELINE COURT E.
Manufacturer CityEARTH CITY MO 63045
Manufacturer CountryUS
Manufacturer Postal63045
Manufacturer Phone3143440010
Manufacturer G1YOUNG DENTAL MANUFACTURING CO 1, LLC
Manufacturer Street13705 SHORELINE COURT E.
Manufacturer CityEARTH CITY MO 63045
Manufacturer CountryUS
Manufacturer Postal Code63045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameYOUNG DENTAL STANDARD POINTED PROPHY BRUSH - 090201
Generic NameSTANDARD POINTED PROPHY BRUSH
Product CodeEHK
Date Received2016-05-09
Model Number090201
Catalog Number090201
Lot NumberUNKNOWN
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerYOUNG DENTAL MANUFACTURING CO 1, LLC
Manufacturer Address13705 SHORELINE COURT E. EARTH CITY MO 63045 US 63045


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-05-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.