APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2016-05-10 for APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR 4407205 manufactured by Life Technologies Holdings Pte Ltd..

Event Text Entries

[44891971] On (b)(4) 2016 a (b)(4) scientific field service engineer visited the (b)(6) to respond to the user report and service the 7500 fast dx instrument after the incident. The engineer noted the laptop computer was on located on a holder on top of the 7500 fast dx instrument. The holder was made of metal with rubber feet and rubber pads. The laptop mouse was placed on a mousepad beside the 7500 fast dx instrument. The instrument and mousepad were located on a lab bench made of metal with a plastic resin top. The engineer visually inspected the 7500 fast dx and noticed four (4) loose screws in back of the instrument. It appeared that the loose screws caused the instrument cover to become loose. The cover was removed from the instrument and grounding cables inside the instrument were visually inspected for attachment to correct sources. All grounding cables inside the instrument were attached correctly. The engineer then tightly screwed in the cover to close the instrument. The laptop and instrument were both plugged into a cyber power ups / surge protector. The engineer checked the electrical outlet for proper grounding using a line monitoring kit. The outlet did not have any defects. The power cord for the laptop was replaced and the instrument and laptop were turned on. Both the instrument and laptop were operating correctly. Self-tests on the 7500 fast dx system were run. The instrument passed specifications. No visual signs of electrical damage on the instrument or laptop were observed. No signs of software corruption from electrical damage were observed. On (b)(4) 2016 a representative from (b)(4) contacted the reporter, (b)(6), to inquire about adverse health effects from the incident. The user reported no adverse health issues and stated the facility is waiting to install a static mat, or other safety measure, before using the instrument again.
Patient Sequence No: 1, Text Type: N, H10


[44891972] A user reported a "shock" while wearing gloves that felt like a surge and vibration. The user had one hand on the computer mouse and one hand on the computer keyboard with the 7500 fast dx instrument turned on. The sensation lasted a few seconds and caused the user to feel tired and a slight tingling sensation in their forehead. The user reported recovering from the initial shock quickly. They were sent to the emergency room for precaution and were released by the emergency room doctor with no further hospitalization. No adverse reactions were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003673482-2016-00003
MDR Report Key5643862
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2016-05-10
Date of Report2016-04-11
Date of Event2016-04-01
Date Mfgr Received2016-04-11
Device Manufacturer Date2014-07-09
Date Added to Maude2016-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKELLI TANZELLA
Manufacturer Street3175 STALEY ROAD
Manufacturer CityGRAND ISLAND NY 14072
Manufacturer CountryUS
Manufacturer Postal14072
Manufacturer Phone7167743122
Manufacturer G1LIFE TECHNOLOGIES HOLDINGS PTE LTD.
Manufacturer StreetBLK 33, MARSILING INDUSTRIAL ESTATE ROAD 3, NO. 07-06
Manufacturer CitySINGAPORE NORTH WEST, SINGAPORE 739256
Manufacturer CountrySN
Manufacturer Postal Code739256
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Generic NameAPPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Product CodeNSU
Date Received2016-05-10
Catalog Number4407205
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES HOLDINGS PTE LTD.
Manufacturer AddressBLK 33, MARSILING INDUSTRIAL ESTATE ROAD 3, NO. 07-06 SINGAPORE NORTH WEST, SINGAPORE 739256 SN 739256


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.