RADIAL JAW M00515190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-11 for RADIAL JAW M00515190 manufactured by Boston Scientific.

Event Text Entries

[44930644]
Patient Sequence No: 1, Text Type: N, H10


[44930725] Final biopsy with needle forceps being performed. Physician being assisted by rrt with opening and closing of the forceps. Previous biopsies were uneventful. It appeared that a minute fragment or single tooth of the forceps separated off the main forceps and was in the patient's airway. Another pulmonologist was called for assistance and the fragment was removed with the help of a rat-tooth forceps. However, the fragment was not visualized with the naked eye upon removal. It was likely covered with blood/fibrin and hence difficult to see. Subsequent airway survey did not reveal any foreign body. Stat x-ray of the chest performed which revealed no metallic foreign body in the lung.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5646289
MDR Report Key5646289
Date Received2016-05-11
Date of Report2016-03-01
Date of Event2016-03-09
Report Date2016-03-01
Date Reported to FDA2016-03-01
Date Reported to Mfgr2016-03-01
Date Added to Maude2016-05-11
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRADIAL JAW
Generic NameFORCEPS, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBWH
Date Received2016-05-11
Catalog NumberM00515190
Lot Number18087127
ID Number90402821-06
Device Expiration Date2018-06-25
Device Availability*
Device Age1 DY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-11