VITEK? 2 AST-YS07 TEST KIT 414967

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-05-11 for VITEK? 2 AST-YS07 TEST KIT 414967 manufactured by Biomerieux, Inc.

Event Text Entries

[45031586] A customer in italy notified biomerieux of a discrepant result associated with vitek 2 ast-ys07 test kit. The customer reported the vitek 2 ast-ys07 card found fluconazole to be susceptible for a patient strain while their reference method, sensititre, gave them resistant. The patient strain has been requested from the customer. An internal biomerieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


[48995525] A customer reported false susceptible results for voriconazole (vrc) and fluconazole (flu) on vitek 2 cards (ast-ys07), for a strain of candida albicans, while the customer reference method (sensititre), gave a resistant result. Customer results: strain was subcultured on cba and sda media. Ast-ys07 - lot#- 287346720tested on vitek 2 v7. 01: voriconazole (vrc) minimum inhibitory concentration (mic): <=0. 125mg/l
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1950204-2016-00053
MDR Report Key5646329
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-05-11
Date of Report2016-04-26
Date Mfgr Received2016-04-26
Date Added to Maude2016-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN NONNEMACHER
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196203396
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 AST-YS07 TEST KIT
Generic NameVITEK? 2 AST-YS07 TEST KIT
Product CodeNGZ
Date Received2016-05-11
Catalog Number414967
Lot Number287346720
Device Expiration Date2016-06-11
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC
Manufacturer Address595 ANGLUM ROAD ST. LOUIS MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-11

© 2026 FDA.report