STRATUS CS 10444834

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-11 for STRATUS CS 10444834 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[45160416] Siemens has requested additional information from customer to further evaluate the event. The cause for the event is unknown at this moment.
Patient Sequence No: 1, Text Type: N, H10


[45160417] Customer reported discrepancy between troponin results on stratus cs analyzer. There was no report of injury due to the event.
Patient Sequence No: 1, Text Type: D, B5


[47821819] Message log was not available for investigation as instrument was turned off after the event. Qc is passing and complaint preparer has confirmed that customer is familiar with proper sample handling. Numerous factors can contribute to a discrepant ctni result, including the instrument, reagent testpak, and operators. The cause of the discrepancy in results cannot be determined based up on the information that customer provided. Customer has not had any further issues. Instrument is operational
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2016-00054
MDR Report Key5646715
Date Received2016-05-11
Date of Report2016-06-15
Date of Event2016-04-18
Date Mfgr Received2016-06-10
Date Added to Maude2016-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY
Manufacturer CityNEWARK DE 19702
Manufacturer CountryUS
Manufacturer Postal Code19702
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRATUS CS
Generic NameSTRATUS CS
Product CodeDAP
Date Received2016-05-11
Catalog Number10444834
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-11

© 2024 FDA.report