COBAS 6000 E601 MODULE 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-05-11 for COBAS 6000 E601 MODULE 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[45050192] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[45050193] The customer complained of questionable results for one blood sample while testing for elecsys ferritin (ferritin). It was stated that they received a result of 0. 515 ng/ml without any flags on an aliquoted tube on cell 1. They reran the sample and got 695. 1 ng/ml without any flags on the same aliquoted sample on the same cell. They reran the primary tube on cell 2 and obtained a value of 728 ng/ml. All of these results were from the same analyzer. They originally reported the 0. 515 ng/ml result but they "caught it almost immediately and reran it and reported the 695. 1 ng/ml. They believed the 695. 1 value to be the correct result. " there was no adverse event. The reagent is lot number 18714303 with an expiration date of 10/31/2016. The field service representative visit found no cause for the results. A general reagent issue cannot be seen based on the provided quality control data. The instrument was checked and successful performance testing was done. A specific root cause could not be identified. Additional information for further investigation was requested but was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2016-00626
MDR Report Key5648409
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-05-11
Date of Report2016-05-11
Date of Event2016-04-23
Date Mfgr Received2016-04-25
Date Added to Maude2016-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJMG
Date Received2016-05-11
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2016-05-11
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457