MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-05-11 for COBAS 6000 E601 MODULE 04745922001 manufactured by Roche Diagnostics.
[45050192]
(b)(4).
Patient Sequence No: 1, Text Type: N, H10
[45050193]
The customer complained of questionable results for one blood sample while testing for elecsys ferritin (ferritin). It was stated that they received a result of 0. 515 ng/ml without any flags on an aliquoted tube on cell 1. They reran the sample and got 695. 1 ng/ml without any flags on the same aliquoted sample on the same cell. They reran the primary tube on cell 2 and obtained a value of 728 ng/ml. All of these results were from the same analyzer. They originally reported the 0. 515 ng/ml result but they "caught it almost immediately and reran it and reported the 695. 1 ng/ml. They believed the 695. 1 value to be the correct result. " there was no adverse event. The reagent is lot number 18714303 with an expiration date of 10/31/2016. The field service representative visit found no cause for the results. A general reagent issue cannot be seen based on the provided quality control data. The instrument was checked and successful performance testing was done. A specific root cause could not be identified. Additional information for further investigation was requested but was not provided.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2016-00626 |
MDR Report Key | 5648409 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2016-05-11 |
Date of Report | 2016-05-11 |
Date of Event | 2016-04-23 |
Date Mfgr Received | 2016-04-25 |
Date Added to Maude | 2016-05-11 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS 6000 E601 MODULE |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | JMG |
Date Received | 2016-05-11 |
Model Number | E601 |
Catalog Number | 04745922001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS 6000 E601 MODULE |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2016-05-11 |
Model Number | E601 |
Catalog Number | 04745922001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-05-11 |