MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-05-13 for I-STAT B-HCG CARTRIDGE 05P58-25 manufactured by Abbott Point Of Care.
[45197399]
(b)(4).
Patient Sequence No: 1, Text Type: N, H10
[45197400]
On 04/28/2016, abbott point of care (apoc) was contacted by a customer regarding i-stat b-hcg cartridges that yielded a suspected false positive result on a (b)(6) year old female patient admitted with back pain and fever. There was no additional patient information at the time of this report. The patient was given a transvaginal ultrasound, which confirmed the patient was not pregnant. The customer states that return product is available for investigation. Date: (b)(6) 2016, time: 17:00, i-stat: 34, lab: <1. Based on the information available and internal control algorithms the event is reportable as a potential product malfunction although not confirmed at this time. There are no injuries associated with this event. The investigation is underway.
Patient Sequence No: 1, Text Type: D, B5
[49437061]
Apoc incident # (b)(4). The investigation was completed on 06/22/2016. Retain product was tested and is functioning according to specification. Return product was not available for investigation.
Patient Sequence No: 1, Text Type: N, H10
[49437062]
Na
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2245578-2016-00025 |
MDR Report Key | 5652416 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2016-05-13 |
Date of Report | 2016-07-11 |
Date of Event | 2016-03-28 |
Date Mfgr Received | 2016-06-22 |
Device Manufacturer Date | 2015-12-03 |
Date Added to Maude | 2016-05-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LINDA MACZUSZENKO |
Manufacturer Street | 400 COLLEGE ROAD |
Manufacturer City | PRINCETON NJ 08540 |
Manufacturer Country | US |
Manufacturer Postal | 08540 |
Manufacturer Phone | 6136885949 |
Manufacturer G1 | ABBOTT POINT OF CARE CANADA LTD. |
Manufacturer Street | 185 CORKSTOWN ROAD |
Manufacturer City | OTTAWA, ONTARIO K2H 8V4 |
Manufacturer Country | CA |
Manufacturer Postal Code | K2H 8V4 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | I-STAT B-HCG CARTRIDGE |
Generic Name | B-HCG CARTRIDGE |
Product Code | NAL |
Date Received | 2016-05-13 |
Model Number | NA |
Catalog Number | 05P58-25 |
Lot Number | A15337 |
Device Expiration Date | 2016-05-14 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT POINT OF CARE |
Manufacturer Address | 400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-05-13 |