MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-05-13 for SIMPLEXA FLU A/B & RSV DIRECT AND SIMPLEXA FLU A/B & RSV POSITIVE CONTROL MOL2650 NA manufactured by Focus Diagnostics.
[45202489]
Focus diagnostics has received a total of (b)(4) complaints from 21 customers for early cycle threshold (ct) issues with the direct amplification disc (dad) (mol1455, mol1451, and mol 1452) while performing the simplexa flu a/b & rsv direct assay. This error may result in a false positive result, a false negative result, and error code(s) giving an invalid run. Early internal testing reproduced the early ct threshold; however, these early ct were produced only when the dad was used after a previous partial run. The false result was not reported to the doctor. There was no patient involvement. A voluntary recall has been performed related to the reported lot. As of may 13, 2016, there has been no report of patient injury or death.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2023365-2016-00042 |
MDR Report Key | 5654999 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2016-05-13 |
Date of Report | 2016-05-13 |
Date of Event | 2016-04-01 |
Date Mfgr Received | 2016-04-01 |
Date Added to Maude | 2016-05-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS VALERIE CIMMARUSTI |
Manufacturer Street | 11331 VALLEY VIEW STREET |
Manufacturer City | CYPRESS CA 90630 |
Manufacturer Country | US |
Manufacturer Postal | 90630 |
Manufacturer Phone | 5622406484 |
Manufacturer G1 | FOCUS DIAGNOSTICS |
Manufacturer Street | 11331 VALLEY VIEW STREET |
Manufacturer City | CYPRESS CA 90630 |
Manufacturer Country | US |
Manufacturer Postal Code | 90630 |
Single Use | 3 |
Remedial Action | NO |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SIMPLEXA FLU A/B & RSV DIRECT AND SIMPLEXA FLU A/B & RSV POSITIVE CONTROL |
Generic Name | RESPIRATORY VIRUS PANEL NUCLEIC ACID ASSAY SYSTEM |
Product Code | OCC |
Date Received | 2016-05-13 |
Model Number | MOL2650 |
Catalog Number | NA |
Lot Number | 29763 |
Operator | OTHER HEALTH CARE PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | FOCUS DIAGNOSTICS |
Manufacturer Address | 11331 VALLEY VIEW STREET CYPRESS CA 90630 US 90630 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2016-05-13 |