PROGRAMMING SOFTWARE MODEL 250 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-16 for PROGRAMMING SOFTWARE MODEL 250 NA manufactured by Cyberonics - Houston.

Event Text Entries

[46511596]
Patient Sequence No: 1, Text Type: N, H10


[46511597] A handheld device (hhd) was received with an ac adapter, serial cable, flashcard software, and programming wand. Upon receipt, the main battery was depleted. The handheld device was powered using the returned ac power supply with no anomalies. During startup, the charge light on the power button was flashing green and amber indicating that the main battery was not being charged reliably by the handheld. The hhd was opened and a visual inspection identified that the solder connections between the main battery terminal and main board were broken. Once the connections were re-soldered, battery charging returned to normal and the handheld device functioned normally. The handheld was powered-on continuously using only the main battery for more than an hour. At the conclusion of testing, the main battery had a remaining charge of 42%. No other anomalies were identified. No anomalies associated with flashcard software or databases were identified during the flashcard analysis. The flashcard and software performed according to functional specifications. No anomalies were found with the returned programming wand.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644487-2016-01024
MDR Report Key5656619
Date Received2016-05-16
Date of Report2016-05-09
Date of Event2016-05-09
Date Mfgr Received2016-05-09
Device Manufacturer Date2008-01-31
Date Added to Maude2016-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. NJEMILE CRAWLEY
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal77058
Manufacturer Phone2812287200
Manufacturer G1CYBERONICS - HOUSTON
Manufacturer Street100 CYBERONICS BLVD SUITE 600
Manufacturer CityHOUSTON TX 77058
Manufacturer CountryUS
Manufacturer Postal Code77058
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NamePROGRAMMING SOFTWARE
Generic NamePROGRAMMING COMPUTER
Product CodeLJY
Date Received2016-05-16
Returned To Mfg2016-04-18
Model NumberMODEL 250
Catalog NumberNA
Lot Number841226
ID NumberVERSION 8.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCYBERONICS - HOUSTON
Manufacturer Address100 CYBERONICS BLVD NA HOUSTON TX 77058 US 77058


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-16

© 2026 FDA.report