ARCHITECT STAT CK-MB 02K42-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-16 for ARCHITECT STAT CK-MB 02K42-25 manufactured by Abbott Laboratories.

Event Text Entries

[45600055] An evaluation is in process. A follow up report will be submitted when the evaluation is complete. (b)(4). An evaluation is in process.
Patient Sequence No: 1, Text Type: N, H10


[45600056] The customer observed a falsely elevated ck-mb result for one patient on the architect i1000sr analyzer. The following data was provided sid (b)(4): initial 57. 5, repeat 2. 0, 1. 9 (new sample). There was no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[48206243] Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing. No adverse trend was identified for the customer's issue. Labeling was reviewed and found to be adequate. The product was not available for return. An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications. Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[73921078] The device evaluation was reassessed and concluded that a malfunction occurred, therefore, the device was not performing as intended, however, no systemic issue or product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2016-00053
MDR Report Key5657921
Date Received2016-05-16
Date of Report2017-04-14
Date of Event2016-04-27
Date Mfgr Received2017-04-10
Device Manufacturer Date2015-09-01
Date Added to Maude2016-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600643500
Manufacturer CountryUS
Manufacturer Postal Code600643500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT STAT CK-MB
Generic NameCK-MB
Product CodeJHX
Date Received2016-05-16
Catalog Number02K42-25
Lot Number31595UN15
Device Expiration Date2017-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.