DIMENSION EXL WITH LM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-05-18 for DIMENSION EXL WITH LM manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[45478352] A siemens customer service engineer (cse) was dispatched to the customer site. The cse changed the sample drain, two sample pump valves and the vessel handling. The cause of the discordant, falsely elevated tacrolimus result is unknown. No further issues were found since the cse had visited the customer. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[45478353] A discordant, falsely elevated tacrolimus result was obtained on a patient sample on a dimension exl with lm instrument. The discordant result was not reported to the physician(s). The initial result was obtained without dilution. The sample was repeated with dilution (1:2) on the same instrument, resulting lower. The customer reran the sample again without dilution resulting lower than the initial result. There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated tacrolimus results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1226181-2016-00291
MDR Report Key5665346
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-05-18
Date of Report2016-05-18
Date of Event2016-04-20
Date Mfgr Received2016-04-21
Device Manufacturer Date2010-03-09
Date Added to Maude2016-05-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MARGARITA KARAN
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION EXL WITH LM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeMLM
Date Received2016-05-18
Model NumberDIMENSION EXL WITH LM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.