SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2016-05-19 for SYSTEM, PERITONEAL, AUTOMATIC DELIVERY manufactured by Baxter Healthcare - Mountain Home.

Event Text Entries

[45574980] (b)(4). This is a report of use error where a patient replaced a solution bag during therapy. Use errors and proper user instructions are addressed in? The homechoice and homechoice pro systems patient at-home guide? , which is shipped with every homechoice device. The guide instructs the user not to attempt to reuse any disposable supplies. It warns the user not to replace empty solution bags or reconnect disconnected solution bags during therapy. Also, it warns the user that possible contamination of the fluid or fluid pathways can result if disposables are reused. A review of the label for the product family will be conducted. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[45574981] It was reported that a patient reused single-use supplies during fill three of four of peritoneal dialysis therapy. The patient was connected at the time of the event. The patient replaced the heater bag during therapy. The technical service representative assisted with ending therapy. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2016-09438
MDR Report Key5667984
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2016-05-19
Date of Report2016-05-19
Date of Event2016-04-25
Date Mfgr Received2016-04-25
Date Added to Maude2016-05-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKINGA ALMASAN
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Street1900 N HIGHWAY 201
Manufacturer CityMOUNTAIN HOME AR 72653
Manufacturer CountryUS
Manufacturer Postal Code72653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Generic NameSYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Product CodeFKK
Date Received2016-05-19
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Address1900 N HIGHWAY 201 MOUNTAIN HOME AR 72653 US 72653


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-19