STABLELOC EXTERNAL FIXATOR KIT FX-4001-S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-19 for STABLELOC EXTERNAL FIXATOR KIT FX-4001-S manufactured by Acumed Llc.

Event Text Entries

[45576463] Evaluation of the returned product found that there was a driver tip broken off in the set screw. (this occurred during the removal surgery when it broke trying to loosen the set screw. ) the set screw was unscrewed by hand - there was no evidence of binding between the screw and the housing. There was also no evidence of cross threading or deformation. All of the parts of the returned device were found to be in spec.
Patient Sequence No: 1, Text Type: N, H10


[45576464] The surgeon put a stableloc external fixator on a comminuted distal radius fracture on (b)(6) 2016. On (b)(6) 2016, the patient was seen for removal. The surgeon could not loosen the set screw to remove it; a driver tip was broken off in the screw head. The patient was brought to the surgery center for removal the next day and was placed under anesthesia. The surgeon used pliers to loosen the screw and remove the product.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3025141-2016-00127
MDR Report Key5668058
Date Received2016-05-19
Date of Report2016-05-09
Date of Event2016-04-27
Date Mfgr Received2016-05-09
Device Manufacturer Date2013-11-11
Date Added to Maude2016-05-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MICKI LEHMAN
Manufacturer Street5885 NW CORNELIUS PASS ROAD
Manufacturer CityHILLSBORO OR 971249432
Manufacturer CountryUS
Manufacturer Postal971249432
Manufacturer Phone5036279957
Manufacturer G1ACUMED LLC
Manufacturer Street5885 NW CORNELIUS PASS ROAD
Manufacturer CityHILLSBORO OR 971249432
Manufacturer CountryUS
Manufacturer Postal Code971249432
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTABLELOC EXTERNAL FIXATOR KIT
Generic NameCOMPONENT, TRACTION, INVASIVE
Product CodeJEC
Date Received2016-05-19
Returned To Mfg2016-05-13
Model NumberFX-4001-S
Catalog NumberFX-4001-S
Lot Number313110
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerACUMED LLC
Manufacturer Address5885 NW CORNELIUS PASS ROAD HILLSBORO OR 971249432 US 971249432


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-05-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.