SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2016-05-20 for SYSTEM, PERITONEAL, AUTOMATIC DELIVERY manufactured by Baxter Healthcare - Mountain Home.

Event Text Entries

[45576520] (b)(4). An alarm indicative of a potential malfunction of the disposable cassette was reported. The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[45576521] It was reported that a homechoice device experienced a system error 2240 (air in line/set) alarm. The patient was connected at the time of the alarm. This occurred during drain five of five of peritoneal dialysis therapy. There was nothing unusual found during troubleshooting that would cause or contribute to the alarm. The technical service representative assisted with ending therapy. There was no patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2016-09477
MDR Report Key5669293
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2016-05-20
Date of Report2016-05-20
Date of Event2016-04-27
Date Mfgr Received2016-04-27
Date Added to Maude2016-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKINGA ALMASAN
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Street1900 N HIGHWAY 201
Manufacturer CityMOUNTAIN HOME AR 72653
Manufacturer CountryUS
Manufacturer Postal Code72653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Generic NameSYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Product CodeFKK
Date Received2016-05-20
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - MOUNTAIN HOME
Manufacturer Address1900 N HIGHWAY 201 MOUNTAIN HOME AR 72653 US 72653


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-20

© 2026 FDA.report