PETIT-POINT MIXTER FCPS 7-1/4 140223

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-23 for PETIT-POINT MIXTER FCPS 7-1/4 140223 manufactured by Hebumedical Gmbh.

Event Text Entries

[46857896] Per complaint files of the last three years, no similar issues were recorded for this device. The probable root cause is inappropriate use. As per event description the clamp was used for a traction and reduction maneuver during a percutaneous pinning operation. The petit-point-mixter fcps 7-1/4 is intended for use as a vascular clamp. Its delicate tips are not designed for traction maneuvers with high tension forces.
Patient Sequence No: 1, Text Type: N, H10


[46857897] As per importer report # (b)(4): a piece of the instrument broke off in the patient's foot during surgery. (b)(6) 2015 customer reported that on (b)(6) 2015 a left foot percutaneous pinning of numbers 3,4 and 5 metatarsals was performed. During a traction and reduction maneuver the tip of one of the arms of the clamp broke off and lodged deep in the soft tissue. Multiple attempts made with fluoroscopy, ap lateral and obliques were fruitless to find broken piece, given the size of the foot due to swelling and the depth of the incision. Post op visits on (b)(6) 2015 indicate progressive healing. (b)(6) 2015 x-ray shows small metal fragment. Healing as expected, no issues related to the metal fragment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002806609-2016-00009
MDR Report Key5672309
Date Received2016-05-23
Date of Report2016-05-20
Date of Event2015-02-18
Date Mfgr Received2015-12-08
Date Added to Maude2016-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. THOMAS BUTSCH
Manufacturer StreetBADSTRASSE 8
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal78532
Manufacturer Phone46194710
Manufacturer G1HEBUMEDICAL GMBH
Manufacturer StreetBADSTRASSE 8
Manufacturer CityTUTTLINGEN, 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePETIT-POINT MIXTER FCPS 7-1/4
Generic NameHEMOSTAT
Product CodeHRQ
Date Received2016-05-23
Catalog Number140223
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHEBUMEDICAL GMBH
Manufacturer AddressBADSTRASSE 8 TUTTLINGEN, 78532 GM 78532


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-05-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.