RECOVERY CONE REMOVAL SYSTEM RC15

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-23 for RECOVERY CONE REMOVAL SYSTEM RC15 manufactured by Bard Peripheral Vascular, Inc..

Event Text Entries

[46200906] No medical records or no images have been made available to the manufacturer. As the lot number for the device was provided, a review of the device history records is currently being performed. The device has been returned to the manufacturer for evaluation. The investigation of the reported event is currently underway. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[46200907] It was reported that during a vena cava filter retrieval procedure, the marker band allegedly detached from the sheath of the vena cava filter removal system. The filter and the detached marker band were successfully captured and retrieved. There was no reported impact or consequence to the patient.
Patient Sequence No: 1, Text Type: D, B5


[50765399] Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing. This lot met all release criteria. There was nothing found to indicate there was a manufacturing related cause for this event. This is the only event reported to date for this lot number and failure mode. Visual/microscopic inspection: the cone was received inserted through the retrieval sheath. The distal tip of the sheath was buckled. The distal marker band was returned detached from the sheath. The distal end of the introducer sheath was examined under microscopic magnification. The distal marker band impression was identified on the appropriate location on the sheaths' surface. This indicates that the marker band was swaged on the introducer sheath during manufacturing. No other anomalies were identified on the sheath. Medical records review: as medical records were not provided, a review could not be performed. Image/photo review: no images or photos have been made available to the manufacturer. Conclusion: the recovery cone removal system was returned. The distal tip of the introducer sheath was returned buckled and the marker band was returned detached from the catheter. The investigation is confirmed for a detached marker band and buckled material. Based upon the available information, the definitive root cause is unknown. Labeling review: the current ifu (instructions for use) states: general information: the recovery cone removal system is intended to percutaneously remove the g2 x filter, g2 express filter, g2 filter, recovery filter or a foreign body as indicated. Indications for use: the recovery cone removal system is intended for use to percutaneously remove the g2 x filter, g2 express filter, g2 filter and the recovery filter from the vena cava, or facilitate the retrieval of foreign objects from the peripheral vascular system. Warnings: do not use excessive force when manipulating the cone. Excessive force may damage the catheter or other parts of the recovery cone removal system. Equipment required: 12 french dilator, directions for use - g2 x filter, g2 express filter, g2 filter or recovery filter removal.. Insert the guidewire and gently advance it to the location of the g2 x filter, g2 express filter, g2 filter or recovery filter for removal. Pre-dilate the accessed vessel with a 12 french dilator. Advance the 10 french introducer catheter together with its tapered dilator over the guidewire and into the vein, such that the tip of the sheath is approximately 3cm cephalad to the filter tip. Note: the introducer catheter has a radiopaque marker at the distal end of the catheter sheath to assist in visualization. Perform a standard inferior venacavogram (typically 30 ml of contrast medium at 15 ml/s). Check for thrombus within the filter. If there is significant thrombus within the filter, do not remove the g2 x filter, g2 express filter, g2 filter or recovery filter. Capture and removal of the g2 x filter, g2 express filter, g2 filter or the recovery filter: the capture of the g2 x filter, g2 express filter, g2 filter or recovery filter is illustrated in - figure 4 a-e: (b)(4). The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
Patient Sequence No: 1, Text Type: N, H10


[50765400] It was reported that during a vena cava filter retrieval procedure, the marker band allegedly detached from the sheath of the vena cava filter removal system. The filter and the detached marker band were successfully captured and retrieved. There was no reported impact or consequence to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020394-2016-00460
MDR Report Key5673774
Date Received2016-05-23
Date of Report2016-04-28
Date of Event2016-04-28
Date Mfgr Received2016-07-27
Device Manufacturer Date2015-08-26
Date Added to Maude2016-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1625 W 3RD ST.
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1C.R. BARD, INC. (GFO)
Manufacturer Street289 BAY ROAD
Manufacturer CityQUEENSBURY NY 12804
Manufacturer CountryUS
Manufacturer Postal Code12804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRECOVERY CONE REMOVAL SYSTEM
Generic NameVENA CAVA FILTER REMOVAL SYSTEM
Product CodeGAE
Date Received2016-05-23
Returned To Mfg2016-05-09
Catalog NumberRC15
Lot NumberGFZG3634
Device Expiration Date2018-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBARD PERIPHERAL VASCULAR, INC.
Manufacturer Address1625 W 3RD ST. TEMPE AZ 85281 US 85281